SpiroWay Pro - User Manual

SpirWay pro - Setup 1 SpiroWay Pro - Setup 2

ExclamationSpiroWay Pro This product is intended to be used only with SpiroClinic Pro. You should replace this product as often as the manufacturer specifies.

SpiroWay Pro - Clean- 1

 

SpiroWay Pro - Clean- 2

 

SpiroWay Pro - Clean- 3

 

SpiroWay Pro - Clean- 4

 

Detach and dispose of any BVF mouthpieces connected to the SpiroWay Pro and then remove the SpiroWay Pro from the SpiroClinic Pro handpiece.
To clean the SpiroWay Pro after each patient and whenever visibly soiled;

  • Add dishwashing detergent (e.g. those containing 5-15% anionic surfactant, 5% nonionic surfactant) to warm water to create a soapy solution.
  • Shake the SpiroWay Pro gently in the soapy solution.
  • Hold the SpiroWay Pro under running tap water to rinse
  • Leave the upright on a clean lint-free cloth at room temperature until it is completely dry.

CAUTION: Do not insert the SpiroWay Pro into the SpiroClinic Pro device until it is completely dry. The SpiroWay Pro Airway should be replaced every 3 months. The SpiroWay Pro Airway must be replaced if the patient has or suspected to have a bacterial or viral infection Replace the SpiroWay Pro whenever a risk of cross-contamination is suspected.